The VISX STAR S4 Excimer Laser System



The VISX STAR platform's heritage of dependability and performance is well known. The STAR S4 from VISX has taken that reputation to an even higher level. Our latest excimer laser system represents the most advanced laser technology platform available today. Integrating data collected by the WaveScan WaveFront System, the STAR S4 utilizes these exclusive VISX technologies to deliver precision custom ablations:

  • Variable Spot Scanning (VSS): Variable beam sizes from as small as 0.65 mm up to 6.5 mm scanned over the treatment area, conserving tissue and optimizing treatment times
  • Variable Repetition Rate (VRR): VRR delivers treatment at varying repetition rates, optimizing treatment time
  • ActiveTrak 3-D Active Eye Tracking: Captures all 3 dimensions of intra-operative eye movements—no dilation required
  • ActiveTrak Automatic Centering: Locates, and then automatically sets the treatment center to the ­center of the pupil.

For more detailed information about how the VISX STAR S4 laser integrates with the WaveScan click here.

Room Requirements

Minimum Requirement

  • 10 ft by 13 ft (3.1 m by 4 m) Door opening: 36 in (92 cm)

Recommended

  • 12 ft by 15 ft (3.6 m by 4.5 m) or larger
  • Ambient temperature: 60 to 80° F (15 to 27° C)
  • Humidity requirements: Relative humidity between 35%-65% (non-condensing)
  • Barometric pressure between 11 to 16 psi (1.6 to 2.3 kPa, 12 to17.35 mmHg)
  • Room should be kept clean; dust-/particle-free by tiling floor; do not install carpet; avoid ceiling tiles that shed particles ; no volatile cleaners

System Diminsions and Weight

  • Height: . . . . . . 59 in/149 cm maximum
  • Width: . . . . . . . 50 in/127 cm overall
  • Length: . . . . . . 80 in/203 cm
  • Weight: . . . . . . Main console: 1600 lb/726 kg
  • Patient chair: . . 400 lb/181 kg
  • Patient chair: . . Perpendicular to system at treatment end

Electrical Requirements

Follow all local codes and standards. Electrical conduit to the room must carry:

  • 220 VAC single phase, plus ground, 60 Hz (U.S. specifications)
  • 230 VAC single phase, plus ground, 50 Hz (European specifications)
  • 30 amp service

Note: Power must be greater than 199 VAC and less than 245 VAC