STAR S4 IR Excimer Laser System
|
Room Requirements
Minimum Requirement
- 10 ft by 13 ft (3.1 m by 4 m) Door opening: 36 in (92 cm)
Recommended
- 12 ft by 15 ft (3.6 m by 4.5 m) or larger
- Ambient temperature: 60 to 80° F (15 to 27° C)
- Humidity requirements: Relative humidity between 35%-65% (non-condensing)
- Barometric pressure between 11 to 16 psi (1.6 to 2.3 kPa, 12 to17.35 mmHg)
- Room should be kept clean; dust-/particle-free by tiling floor; do not install carpet; avoid ceiling tiles that shed particles ; no volatile cleaners
System Diminsions and Weight
- Height: . . . . . . 59 in/149 cm maximum
- Width: . . . . . . . 50 in/127 cm overall
- Length: . . . . . . 80 in/203 cm
- Weight: . . . . . . Main console: 1600 lb/726 kg
- Patient chair: . . 400 lb/181 kg
- Patient chair: . . Perpendicular to system at treatment end
Electrical Requirements
Follow all local codes and standards. Electrical conduit to the room must carry:
- 220 VAC single phase, plus ground, 60 Hz (U.S. specifications)
- 230 VAC single phase, plus ground, 50 Hz (European specifications)
- 30 amp service
Note: Power must be greater than 199 VAC and less than 245 VAC |
|
|