Room Requirements

Minimum Requirement

  • 10 ft by 13 ft (3.1 m by 4 m) Door opening: 36 in (92 cm)

Recommended

  • 12 ft by 15 ft (3.6 m by 4.5 m) or larger
  • Ambient temperature: 60 to 80° F (15 to 27° C)
  • Humidity requirements: Relative humidity between 35%-65% (non-condensing)
  • Barometric pressure between 11 to 16 psi (1.6 to 2.3 kPa, 12 to17.35 mmHg)
  • Room should be kept clean; dust-/particle-free by tiling floor; do not install carpet; avoid ceiling tiles that shed particles ; no volatile cleaners

System Diminsions and Weight

  • Height: . . . . . . 59 in/149 cm maximum
  • Width: . . . . . . . 50 in/127 cm overall
  • Length: . . . . . . 80 in/203 cm
  • Weight: . . . . . . Main console: 1600 lb/726 kg
  • Patient chair: . . 400 lb/181 kg
  • Patient chair: . . Perpendicular to system at treatment end

Electrical Requirements

Follow all local codes and standards. Electrical conduit to the room must carry:

  • 220 VAC single phase, plus ground, 60 Hz (U.S. specifications)
  • 230 VAC single phase, plus ground, 50 Hz (European specifications)
  • 30 amp service

Note: Power must be greater than 199 VAC and less than 245 VAC

Privacy Policy | Terms of Use | Contact Us

Copyright © 2009 Abbott Laboratories Inc. Abbott Park, Illinois, USA

Unless otherwise specified, all product names appearing in this Internet site are trademarks owned by or licensed to Abbott Laboratories, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott Laboratories, except to identify the product or services of the company. CareCredit is a registered trademark of CareCredit LLC, General Electric Co. 2009.09.02-IL1281